50
10.2 Executing Calibration
Perform calibration only after correct installation and thorough warm up. Also, make sure
that nothing is on the pan and ensure conditions free from the influence of vibrations or wind.
10.2.1 Span Calibration With Built-in Weight
(CAUW-D/CAUW/CAUX series only)
“i-CAL” (The balance will be adjusted with the built-in calibration weight.)
1
From the mass display, press the
[CAL]
key once. If the display shows “i-CAL”
(Preset calibration method is “i-CAL”),
jump to Step
5
.
If the display shows another item
(Preset calibration method is not “i-CAL”),
go on to Step
2
.
(Refer to 10.3.1 for preset calibration method)
2
Press the
[CAL]
key repeatedly until
“FUnC.SEL” appears, then, press the
[O/T]
key to display “CAL”.
3
Press the
[O/T]
key again. The display shows
“E CAL”.
4
Press the
[CAL]
key twice. The display shows
“i-CAL”.
5
Press the
[O/T]
key. The motor sound of
weight loading system is heard and the
display changes automatically in this order:
“CAL 2”, “CAL 1”, “CAL 0”, and “CAL
End”.
When the display returns to the mass display,
span calibration is complete.
...
Summary of Contents for CAU Series
Page 1: ...2 ...
Page 5: ...6 ...
Page 117: ...118 MEMO ...
Page 118: ...119 ...
Page 119: ...120 ...