81
1
Press the
[CAL]
key repeatedly from the mass
display until “FUnC.SEL” appears. Press the
[O/T]
key. The display will show “CAL”.
2
Press the
[CAL]
key repeatedly until “Addon:
**” appears. The “**” shows the current setting,
“on” for on, “oF” for off.
3
Press the
[O/T]
key. The display shows “Adon-
on” and after this, pressing the
[CAL]
key
switches the displays between “Adon-on” and
“Adon-oF”. When the current setting is
displayed, the stability mark (
¨
) appears.
Press the
[O/T]
key when the desired setting is
shown, to change the setting.
4
Press the
[POWER]
key to return to“Addon:**”
5
Pressing the
[POWER]
key again returns to the
mass display.
6
When the Add-on Mode is set up, the add-on
mode symbol appears in the mass display.
x
No more than one of the four application modes, Auto Print (13.4), Interval
Timer (13.5), Addon Mode (13.6), Formulation Mode (13.7), can be set ON
at the same time.
x
When Add-on mode is in use, fully-automatic span calibration by PSC (10.3.2)
or Clock-CAL (10.3.3) is not performed. Instead, Weight symbol ( )
keeps blinking when span calibration is necessary. Span calibration can be
performed between sets of measurements.
Summary of Contents for CAU Series
Page 1: ...2 ...
Page 5: ...6 ...
Page 117: ...118 MEMO ...
Page 118: ...119 ...
Page 119: ...120 ...