10.2.4 Calibration Check With External Weights
“E-tESt” (The balance is checked with external calibration weights but not adjusted.)
1
From the mass display, press the [CAL] key
once. If the display shows “E tESt”
(Preset calibration method is “E-tESt”),
jump to Step 5.
If the display shows another item
(Preset calibration method is not “E-tESt”),
go on to Step 2.
(Refer to 10.3.1 for preset calibration method)
2
Press the [CAL] key repeatedly until
“FUnC.SEL” appears, then, press the [O/T]
key to display “CAL”.
3
Press the [O/T] key. The display shows “ECAL”.
4
Press the [CAL] key repeatedly until the
display shows “E tESt”.
5
Press the [O/T] key. The calibration test begins
and the zero display blinks.
6
Recheck that there is nothing on the pan and wait.
7
The set weight value (Refer to 10.3.5) appears,
blinking.
8
Place the weight indicated on the pan. Wait
until the blinking zero display reappears.
(It may take about 30 seconds.)
9
Remove the weight. The display shows the drift
from the previous span calibration for several
seconds. Then, after “tESt End” appears for
several seconds, the display returns to the mass
display, and the calibration test is complete.
Note that the Calibration Test procedure DOES
NOT adjust the balance, it only reports the drift
from the previous span calibration.
...