61
10.3.6 Inputting External Calibration Weight Value for PCAL
The exact value (conventional mass value) of the operator’s calibration weight to be used
for PCAL procedure can be entered.
1
From mass display, press the
[CAL]
key
repeatedly until “SEttinG” appears. Press the
[O/T]
key. “CAL dEF” appears.
2
Press the
[CAL]
key repeatedly until “PCAL
Set” appears and press the
[O/T]
key. One of
these numbers will appear, depending on the
machine model: 50.0000g, 100.0000g,
200.0000g, or 300.0000g. In the upper part of
the display panel, the MENU symbol and the #
symbol appear in order to show number entry
mode. The leftmost numeral blinks.
The blinking digit may be changed.
3
When the
[UNIT]
key is pressed, the value of
the blinking digit increases by 1 at a time.
Press the
[PRINT]
key to move the blinking
digit one place to the right. Enter the desired
value.
4
Press the
[O/T]
key to set the external
calibration weight value for built-in weight
calibration. The display proceeds to “SEt”,
then to “PCAL SEt”.
5
Press the
[POWER]
key again to return to
the mass display.
Calibration Weight Value Range
220D
75g to capacity
CAUW-D series
120D
35g to capacity
320
145g to capacity
220
95g to capacity
Othes series
120
45g to capacity
In the CAUW-D series the conventional mass value may be input for up to
five decimal places.
(Example)
(Example)