Hints on Fully-automatic Span Calibration
x
Fully-automatic span calibration
PSC (Not
with CAUY series) (Refer to 10.3.2 for detail.)
Default setting of PSC is ON. Upon detection of temperature change, span calibration is
automatically performed to maintain accuracy. Weight symbol ( ) blinks prior to fully-
automatic calibration.
How to skip a PSC
Press [POWER] key while Weight symbol (
)
is blinking.
How to switch off PSC
From mass display,
press [CAL] key
repeatedly until “SEttinG” appears.
→
Press [O/T] key
to display “CAL dEF”.
→
Press [CAL] key
repeatedly until
“PSC:on” appears.
→
Press [O/T] key.
“PSC-on” appears.
→
Press [CAL] key
. “PSC-oF” appears.
→
Press [O/T] key
.
Stability mark (
¨
) is added.
→
Press [POWER] key
returns “PSC:oF”.
→
Press [POWER] key
repeatedly to return to the mass display.
When PSC is off, perform span calibration according to
Span Calibration Procedure
when it is necessary.
Clock-CAL
Clock-CAL is a practical alternative that performs regular automatic span calibration
according to time setting (CAUW-D/CAUW series only, up to 3 times a day, refer to 10.3.3).