When the Function Does Not Operate At All:
•
For some notebook PCs, it is possible to exclude the RS-232C
port for energy saving purposes. Set the PC so that the RS-232C
port can be used.
•
Change the settings of COM 1 to 4. Re-start Windows* after the
changes are complete.
•
Verify that the correct RS-232C cable is being used.
Windows* = Windows
®
95, Windows
®
98, Windows
®
Me, Windows
®
2000
When the Function Intermittently Malfunctions:
•
Use a communication speed of 300 bps. Depending on the
processing ability of the PC, this function may operate incorrectly
if communication speed is too high.
•
Send the next data only after the current one is displayed on the
screen. Depending on the processing ability of the PC, this
function may operate incorrectly if interval of data transmission is
too short.
•
Do not touch the keyboard or the mouse while the balance is
transmitting data.
Stop the data transmission and confirm that no data is entering the
PC before touching the keyboard or the mouse.
•
This operation may be incorrect when the displayed value is not a
weight value (i.e. error code or time).
•
The unit designation is not transmitted.
•
This function may operate incorrectly depending on the settings
of various lock keys of the keyboard such as the NUMLOCK or
cursor key lock. Change the state of the lock and function keys on
the PC keyboard.
•
The peripheral devices connected to the DATA I/O such as
Electric Printer EP-50 cannot be used with this function.
•
When this function is used, a command cannot be sent from the
peripheral device to the balance.
•
Set the data formats, such as decimal places and units, in each
application.