33
(3) To change the setting, press the
→
O/T
←
key while
"Addon:**"
is
displayed.
Press the CAL/MENU key to switch between
"Adon-oF"
and
"Adon-
on".
The stability symbol illuminates to indicate the current setting.
(4) Press the
→
O/T
←
key to change the setting, or press the
POWER/BRK key to return to
"Addon:**"
without changing the
settings.
(5) When
"Addon:**"
is displayed, press and hold the POWER/BRK
key to return to the mass display.
Regardless of the unit displayed, 1 count is equivalent to 0.1 mg.
14. STANDARD SPAN CALIBRATION (Sensitivity Adjustment)
There are four types of calibration available for the CAW/CAX series
balances and two for the CAY series balances. These calibration types are
described in Section
7. “SPAN CALIBRATION.”
The operator sets the type
of calibration that will be performed. The set calibration is the standard span
calibration and appears in the Operation Menu.
Setting the Standard Span Calibration:
Procedure
(1) From the mass display, press the CAL/MENU key several times.
When
"SEttinG"
is displayed, press the
→
O/T
←
key.
(2) When
"CAL dEF"
is displayed, press the
→
O/T
←
key.
(3) Press the CAL/MENU key to display the different types of span
calibration. When the desired type of span calibration is displayed,
press the
→
O/T
←
key to set it. This calibration type is now the
standard span calibration and is displayed with the stability symbol.
•
"E CAL":
Span calibration using an external weight
•
"E tESt":
Span check using an external weight
•
"i CAL":
Span calibration using the built- in weight
(CAW/CAX series only)
•
"i tESt":
Span check using the built- in weight
(CAW/CAX series only)
(4) To return to mass display from
"CAL dEF"
, press and hold the
POWER/BRK key.
Executing the Standard Span Calibration:
Procedure
(1) From the mass display, press the CAL/MENU key to display the
standard span calibration type.
(2) Press the
→
O/T
←
key to execute standard span calibration.