45
• Error such as
ERR0x
is
displayed.
• Hardware failure.
→
[S]
•
Err20
is
displayed.
• The numeric value entered is incorrect.
• The user has attempted to release a registered
unit and the software will not allow it.
→
It is displayed when only one unit is
registered or only the ratio measurement unit
is registered.
•
Err24
is
displayed.
• Low battery pack voltage.
→
Charge the battery pack.
• Display such as
t to 10 appeas.
• Command code for adjustment is being entered.
→
Disconnect the power supply cable. Wait
about 10seconds and reconnect. If this
display is left on the balance for more than
24 hours, it may not be possible to perform
correct measurement.
During
measurem
ent
• The data cannot
be transmitted to
or received from
the peripheral
units.
• Communication parame ter is incorrect.
→
F
See Section
35. “INPUT AND
OUTPUT DATA FORMAT.”
•
CAL E2
is
displayed.
• Something is being placed on the pan during
span calibration.
→
Unload it and wait for span calibration to
restart.
•
CAL E3
is
displayed.
• The reference weight loaded during calibration
of the built- in weight is incorrect.
→
F
See Section
17. “CALIBRATION OF THE
BUILT IN WEIGHT.”
During
span
calibration
•
CAL E4
is
displayed.
• The internal parts of the balance require
service.
→
[S]