8400-M022-O1 Rev K
Model 190 Indicator
22
b=
(Break Range)
The break range is a number from 1 to 255 that corresponds to the number of division
changes to break out of the filtering.
Press the
TARE
key to show the current setting for the break range. If the displayed
setting is acceptable, press the
TARE
key to save it.
Otherwise, use the
Fn/
and
UNITS/
keys to enter a new setting (1 to 255) and press the
TARE
key to save it.
NOTE:
Selecting 0 disables this feature.
Sr=
(Sample Rate)
Press the
TARE
key to show the current setting for the sample rate. The setting displayed
is the sample rate in samples per second. If the displayed setting is acceptable, press the
TARE
key to save it.
Otherwise, use the
Fn/
and
UNITS/
keys to enter a new setting (1 to 120) and press the
TARE
key to save it.
UnS=
(Motion Range)
The motion range is the number of divisions of change permitted before indicating unstable.
Press the
TARE
key to view the current setting for the range of motion detection. If the
displayed setting is acceptable, press the
TARE
key to save it.
Otherwise, use the
Fn/
and
UNITS/
keys to enter a new range and press the
TARE
key to save the new setting. Allowable range values are 0 through 99 divisions.
SC=
(Stable Count)
Stable Count is the number of consecutive stable weight readings before indicating stable
weight. This helps filter weight readings for stability when trying to capture stable weight.
Press the
TARE
key to view the current setting for Stable Count. If the displayed setting
is acceptable, press the
TARE
key to save it.
Otherwise, use the
Fn/
and
UNITS/
keys to enter a new setting and press the
TARE
key to save the new setting. Allowable values for the stable count are 1 through 255.