8400-M022-O1 Rev K
Model 190 Indicator
13
Calibration Data Entry
The Model 190 uses a capacitive touch keypad that requires a “finger touch”
to function. The keypad will not operate with other items such as pens,
pencils, or tools.
Figure No. 9
During the indicator setup and calibration process, it will be necessary to enter operational
parameters via the 190 keypad.
•
Pressing the
TARE
key will cause the data entered or displayed to be retained and
the 190 to advance to the next prompt.
•
The functions of numeric keys are replaced by using the
UNITS/
and the
Fn/
keys.
•
The cursor location is identified by the blinking character and can be advanced to the left
to the next position by pressing the
UNITS/
key.
•
Pressing the
Fn/
key will change the blinking character to the next value or setting.
Continue to press this key to "toggle" between the different available values or settings
for the setup parameter.
•
Pressing the
Fn/
key when a setup parameter (not a parameter value or setting) is
displayed, will "backup" to the previous parameter prompt.