Request Scale to Zero Command
The wireless device sends:
<LF>Z<CR>
The 190 will attempt to zero the scale and respond with weight in SMA format. Scale status
will indicate if the attempt to zero was unsuccessful.
Request Scale to Tare Command
The wireless device sends:
<LF>T<CR>
The 190 will attempt to set the tare to current gross weight and respond with displayed
weight in SMA format. Scale status will indicate if the attempt to tare was unsuccessful.
Setting the tare will change the displayed weight to net mode.
Return Tare Weight Command
The wireless device sends:
<LF>M<CR>
The 190 will respond with the tare weight in SMA format.
Clear Scale Tare Weight Command
The wireless device sends:
<LF>C<CR>
The 190 will clear the tare weight and respond with displayed weight in SMA format.
Clearing the tare will change the displayed weight to gross mode.
Repeat Displayed Weight Continuously Command
The wireless device sends:
<LF>R<CR>
The 190 will repeat displayed weight in SMA format continuously until another SMA
command is received.
About Scale First Line Command
The wireless device sends:
<LF>A<CR>
The 190 will respond with:
<LF>SMA:2/1.0<CR>