8400-M022-O1 Rev K
Model 190 Indicator
21
Analog to Digital Filtering
A-d
With
A-d
displayed, press the
TARE
key. The display will change to
dFLt=
. Proceed to
the
dFLt=
parameter.
A-d?
With
A-d?
displayed, press the
TARE
key. The display will change to
no
. Press the
Fn/
key to toggle to
yes
and press the
TARE
key. The display will change to
dFLt=
.
Proceed to the
dFLt=
parameter.
Otherwise, to skip the Analog to Digital Filtering setup, press the
TARE
key to advance to
the
CAL?
. prompt.
dFLt=
(Digital Filtering)
With
dFLt=
displayed, press the
TARE
key to show the current setting. If the setting
displayed is acceptable, press the
TARE
key to save it.
Otherwise, use the
Fn/
key to toggle to a new setting and press the
TARE
key to save
it. Allowable settings are 0, 1, 2, or 3. Note, that if you select 3 (Custom Filtering) two
additional prompts will be displayed.
dFLt=
0
Filter Level = 2, Break Range = 1
1
Filter Level = 6, Break Range = 12, Sample Rate = 2
2
Filter Level = 20, Break Range = 12, Sample Rate = 1
3
CUSTOM FILTERING
NOTE:
The prompts,
F=
(Filter Level) and
b=
(Break Range) will only be displayed if you
selected 3 (Custom Filtering) for the
dFLt=
(Digital Filtering) prompt.
Digital Filtering (
dFLt=
) selections 0, 1, and 2 have fixed factory settings for Filter Level,
Break Range, and Sample Rate.
F=
(Filter Level)
The filter level is a number from 1 to 99 that corresponds to the level of filtering with 1 being
the least and 99 being the greatest. Press the
TARE
key to show the current setting. To
accept the setting displayed, press the
TARE
key again to save it.
Otherwise, use the
Fn/
and
UNITS/
keys to enter a new setting (1 to 99) and press the
TARE
key to save it.