nCPAP Driver Service Manual
2008, Cardinal Health
Page 59 of 81
2. Plug the Control Knob/Selector Switch into connector J11.
3. Plug the Main to Bulkhead cable into connector J12.
4. Plug the Power Cable Assembly into connector J19.
5. Align the Front Panel tabs into the slots in the Bottom Case.
6. Tilt the front panel back, taking care to tuck the wiring above or below the standoffs
on the Divider/Oxygen Sensor Assembly.
7. Plug the Valve Drive Cable into the Bulkhead to Valves Cable.
8. Verify and/or set the dipswitch configuration for normal operation:
Note: Do not replace the Top Case until the Power-On Self Test is complete and the Patient Treatment
Mode Screen (the main screen) appears.
Verify cable
connections
Verify dip switch
configuration
Pin 1
(on underside)
J27
(Spare Valve 2)
J11
J12
J19