nCPAP Driver Service Manual
2008, Cardinal Health
Page 56 of 81
7. Screw the white Luer lock ring to the Oxygen Sensor Filters connected to the
Divider/Oxygen Sensor Assembly.
8. Ensure that all joints between the two (2) Oxygen Sensor Filters and the Oxygen
Sensor Elbow have remained tight.
Connect luer lock to Oxygen
Sensor Filters (7)
Install 2 mounting screws
Torque to 7 in-lbs (5)
Bulkhead to Valves Connector (6)
CN15 to CN21, (2)
CN16 to CN22, (3)
Press on silicone
tube (1)