nCPAP Driver Service Manual
2008, Cardinal Health
Page 77 of 81
17.
ELECTROMAGNETIC COMPATIBILITY
The following information is directed by the Standard IEC 60601-1-2:2001 for Medical
Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard:
Electromagnetic Compatibility - Requirements and Tests.
WARNING
The use of any power cords with the Air
Life
®
nCPAP System Driver other
than those supplied with the driver may result in increased emissions
and/or decreased Immunity of the Air
Life
®
nCPAP System Driver.
Table 2: Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The Air
Life
®
nCPAP System Driver is intended for use in the electromagnetic environment
specified below. The customer or user of the Air
Life
®
nCPAP System Driver should
assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
CISPR 11
RF Emissions
Group 1
The Air
Life
®
nCPAP System Driver uses RF
energy only for its internal function. RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
CISPR 11
RF Emissions
Class B
IEC 61000-3-2
Harmonic
Emissions
Not
Applicable
IEC 61000-3-3
Voltage
Fluctuations,
Flicker Emissions
Not
Applicable
The Air
Life
®
nCPAP System Driver is suitable for
use in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.