nCPAP Driver Service Manual
2008, Cardinal Health
Page 16 of 81
WARNING
The power switch does not disconnect the driver from the main power
supply. This can only be accomplished by disconnecting the AC Adapter
from either the driver or the AC outlet.
WARNING
Remove electrical power before servicing the unit. The power switch on
this unit does not isolate the driver from the AC Adaptor; disconnect the
power supply cable to ensure complete isolation.
Disconnect oxygen and medical air.
Eye protection is required when servicing this machine.
WARNING
There is always a risk when using oxygen that the O2 level in the arterial
plasma will reach an unacceptable level. This driver should always be
used in conjunction with a pulse oximeter to monitor patient status.
Oxygen toxicity can be a concern with this device for the intended
population if it is used for more than a few hours. Risks include:
•
Absorption atelectasis (collapse of lung tissue)
•
Hypercapnia (increased carbon dioxide)
•
Retrolental hyperplasia (which can cause blindness)
Also, Chemotherapy patients given bleomycin appear to be extremely
sensitive to even relatively low levels of oxygen.
WARNING
Use this product only as directed in the product literature to reduce the
possibility of nasal and skin irritation, and necrosis.