nCPAP Driver Service Manual
2008, Cardinal Health
Page 54 of 81
13.5
Install the Divider/Oxygen Sensor Assembly
CAUTION
This assembly contains electronic components that are ESD sensitive.
ESD control and handling measures should be followed while performing
this process.
1.
Attach the silicone tubing to the fitting on the Bottom Case.
2.
Connect the Oxygen Sensor Cable to Remote Sensor Board (PCA-02) at CN14
3.
Position the Divider/Oxygen Sensor Assembly into its mounting slots.
4.
Gently insert the Remote Board Bulkhead Cable connector through the
corresponding slot in the Divider/Oxygen Sensor Assembly.
Connect Oxygen Sensor
Cable (2)
Attach silicone tube (1)