Section 7: Technical Data
Page 80
©2002 Cardiac Science, Inc.
300282-002 Rev. E0
advise you that the Survivalink AED recommends a defibrillation shock.
When a defibrillation shock is advised, you determine if or when to
deliver the shock.
Cardiac Rhythms Used to Test the Rhythm Recognition
Detection System for Survivalink AED Models 9100/9110
Rhythm Class
Models
ECG
Test
Sample
Size
a
a. From Survivalink ECG rhythm databases.
Specifications
Shockable
Rhythm - VF
9100/9110
304
Survivalink AED meets the AAMI
DF39 requirement and AHA rec-
ommendation
b
of Sensitivity >
90%.
b. Automatic External Defibrillators for Public Access Defibrillation: Recommendations for
Specifying and Reporting Arrhythmia Analysis Algorithm Performance, Incorporating New
Waveforms, and Enhancing Safety, American Heart Association (AHA) AED Task Force and
approved by the AHA Science Advisory and Coordinating Committee, 3/18/97, Table 2.
9200/9210
318
Shockable
Rhythm - VT
9100/9110
12
Survivalink AED meets the AAMI
DF39 requirement and AHA rec-
ommendation
b
of Sensitivity >
75%.
9200/9210
53
Non-shockable
rhythm - Normal
Sinus Rhythms
9100/9110
1082
Survivalink AED meets the AAMI
DF39 requirement of Specificity >
95% and AHA recommendation
b
of Specificity > 99%.
9200/9210
1207
Non-shockable
rhythm - Asys-
tole
9100/9110
10
Survivalink AED meets the AAMI
DF39 requirement and the AHA
recommendation
b
of Specificity >
95%.
9200/9210
16
Non-shockable
rhythm - all other
non-shockable
rhythms
9100/9110
1729
Survivalink AED meets the AAMI
DF39 requirement and AHA rec-
ommendation
b
of Specificity >
95%.
9200/9210
2363