Section 3: Setup
Page 28
©2002 Cardiac Science, Inc.
300282-002 Rev. E0
Operating Mode
- is defined as having the battery installed and the lid
open. This is the mode it would be in during an actual rescue situation.
Standby Mode
- is when the battery is installed, but the lid is closed. This
is the normal mode for the Survivalink AED to be in between rescues. In
this mode, the device will conduct its routine self-tests to determine that
it is operating correctly.
Storage Mode
- is when the battery is removed, such as during shipping
or transport. With the battery removed, the Survivalink AED is unable to
perform self-tests or rescues.
Survivalink AED Operating and Standby Conditions
CAUTION:
Temperature/Humidity/Pressure Extremes
Exposing the Survivalink AED with the battery installed to extremes,
outside the operation and standby conditions, will cause the self-tests to
be disabled and could cause the Survivalink AED to function improperly.
Storing the Survivalink AED outside these conditions for 5 consecutive
days will result in a “service required” alert.
Survivalink AED Shipping and Transport Conditions
(for up to 1 week)
Temperature
0°C to 50°C (32°F to 122°F)
Humidity
5% to 95% (non-condensing)
Atmospheric Pressure
57kPa to 170kPa
Temperature
-40°C to 65°C (-40°F to 149°F)
Temperature W/Display
-30°C to 65°C (-22°F to 149°F)
Humidity
5% to 95% (non-condensing)
Atmospheric Pressure
57kPa to 170kPa