300282-002 Rev. E0
©2002 Cardiac Science, Inc.
Page 75
Section 7: Technical Data
Survivalink AED Models 9100/9110 Waveform
AAMI DF-2 (1996), Section 4.3.4.2, truncated exponential
200 Joule Survivalink AED Models 9100/9110 Waveform
(all values are typi
cal)
300 Joule Survivalink AED Models 9100/9110 Waveform
(all values are typical)
Patients’ Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
25
1900
3.0
50
1900
4.0
75
1900
6.1
100
1900
8.1
125
1900
10.1
Patients’ Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
25
2000
3.5
50
2000
6.9
75
2000
10.4
100
2000
13.9
125
2350
9.8
0
500
1000
1500
2000
2500
0
5
10
15
Time (ms)
Output
Voltage
(V)