300282-002 Rev. E0
©2002 Cardiac Science, Inc.
Page 55
Section 6: Maintenance & Troubleshooting
improperly. Storing the Survivalink AED outside these conditions for 5
consecutive days will result in a “service required” alert.
• Temperature
0°C to 50°C (32°F to 122°F)
• Humidity
5% to 95% (non-condensing)
• Pressure
57kPa (+15,000 ft.) to 170kPa (-15,000 ft.)
Daily Self-Test
Every day, at 3:03:03 a.m., the daily self-test occurs; checking the
battery, electrodes, electronics,“Rescue/Resume” button
and the
software.
These components are also tested each time you
open
and
close
the Survivalink AED lid.
Monthly Self-Tests
A
full charge
is discharged internally to test the
high voltage circuitry
.
This additional test occurs every 28 days when the daily self-test occurs.