300282-002 Rev. E0
©2002 Cardiac Science, Inc.
Page 63
Section 6: Maintenance & Troubleshooting
A This message occurs when there is a previously stored rescue in the internal mem-
ory of the Survivalink AED AND:
The Rescue Data Card memory is full; or
The Rescue Data Card is not inserted
You can clear the message by:
1. Inserting a blank Rescue Data Card and transferring the rescue data from the
internal memory of the Survivalink AED to the Rescue data card and erasing
the stored rescue data (do this if the message does not occur during a rescue
attempt)
2. Downloading the rescue data with RescueLink and erasing the stored rescue data
3. Pressing the “Rescue/Resume” button to erase the internally stored rescue data
4. Replacing a full Rescue Data Card with an empty one (do this when the message occurs
during a rescue attempt)
13.
Q When I open the lid, why do I get the voice prompt
“Card full. Storing internally?”
A There is a previous rescue in the optional Rescue Data Card, and the rescue data
from the current rescue will be stored in the internal memory of the Survivalink
AED.
An invalid card (other than the Rescue Data Card) can also cause this prompt.
Questions and Answers