Section 7: Technical Data
Page 76
©2002 Cardiac Science, Inc.
300282-002 Rev. E0
360 Joule Survivalink AED Models 9100/9110 Waveform
(all values are typical)
Energy Levels and Patient Impedance
The Survivalink AED Monophasic Truncated Exponential (MTE)
waveform has energy levels of 200 J, 300 J, 360 J (± 15%) at 50 Ohms
patient impedance. This allows for escalating energies of subsequent
shocks. The automatic waveform correction for patient impedance is
shown in the preceding Survivalink AED Models 9100/9110 Waveform
tables.
Patients’ Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
25
2000
5.8
50
2000
11.5
75
2000
17.3
100
2000
23.0
125
2350
13.2