Section 6: Maintenance & Troubleshooting
Verifying the Integrity of the Service Indicator (LED) and Cir-
cuitry
To verify the integrity of the service indicator (LED) and circuitry, do the
following:
1. Immediately after opening the Survivalink AED lid, press and hold
the “Rescue/Resume” button and confirm that the “Service” LED is
lit.
2. Release the “Rescue/Resume” button.
3. Close the lid.
4. Verify that the “Status” indicator returns to GREEN.
5. Open the lid and confirm that no diagnostic indicators are lit.
6. Close the lid.
Verifying the Integrity of the Case
Examine the molded case of the Survivalink AED for any visible signs of
stress. If the case shows signs of stress, contact Technical Support at one
of the following telephone numbers:
(888) 466-8686 (USA only)
(952) 939-4181 (USA and Canada)
+45 44 38 05 39 (International)
Cleaning the Survivalink AED Case
Gently clean the surface of the Survivalink AED case with a damp
sponge or with a cloth and mild soap.
CAUTION:
Case Cleaning Solutions
When disinfecting the case, use a non-oxidizing disinfectant, such as
ammonium salts or a gluteraldehyde based cleaning solution, to avoid
damage to the metal connectors.
Verifying the Time Clock
At regular intervals, verify the time on the Survivalink AED using
RescueLink. To adjust the internal clock, follow the directions in the
RescueLink User Manual.