Section 7: Technical Data
Page 74
©2002 Cardiac Science, Inc.
300282-002 Rev. E0
Environmental Conditions
Shipping and Transport Conditions
ASTM D4169-92
Condition
Models
Standard or Compliance
Temperature/
Altitude/
Decompres-
sion/Overpres-
sure
9100/9110
9200/9210
RTCA/DO-160D, Section 4, Category
A4, Operating: 0°C to 50°C, Ground Sur-
vival: 0°C to 50°C
Temperature
Variation
9200/9210 RTCA/DO-160D, Section 5, Category C
Free Fall Drop
9100/9110
9200/9210
IEC 68-2-32 (1975), 1 meter
Shock (Bump)
9100/9110
IEC 68-2-29, 25g and 6000 bumps
9200/9210 IEC 68-2-29, 40g and 6000 bumps
Vibration
(Random)
9100/9110
9200/9210
RTCA/DO-160D, Section 8, Category S
9200/9210 IEC 68-2-64: 10Hz - 2kHz at 0.0500 -
0.0012 g2/Hz
Vibration
(Sine)
9100/9110
IEC 68-2-6: 10Hz - 40.7Hz at 0.15 mm
and 40.7Hz - 150Hz at 1g
9200/9210 IEC 68-2-6: 10Hz - 57.6Hz at 0.15 mm
and 57.6Hz - 150Hz at 2g
Enclosure Pro-
tection
9100/9110
9200/9210
IEC 529, IP23