CAPINTEC, INC.
CRC
®
-25R
Entering Serial Number
The serial number may be any combination of 10 numbers and letters.
Co57
ENTER S/N
_
Figure 6-9 Entering Serial Number
To input the serial number of the Test Source, use the NUMBER / LETTER keys. When a
key is pressed, the number appears. Press the key again to change the number into one of
the letters on the same key (e.g. press the
2
key 3 times to get the letter B). To go to the next
character, press another key. If the next character is on the same key, press the
UP ARROW
(
)
to repeat use of that key. The
DOWN ARROW (
)
is the space key
Press
ENTER
when correct serial number has been input. Figure 6-10 Entering Calibration
Date will appear.
Entering Calibration Date
The calibration date is input next.
Co57
CALIBRATED
_
MMDDYYYY
Figure 6-10 Entering Calibration Date
Input the Test Source calibration date in the form MMDDYYYY (example: July 3, 2005 would
be 07032005) and press
ENTER
. The date is checked for validity. If the entered date is not
valid, a beep will sound and “DATE ERROR” will be displayed. The date must be re-entered.
Figure 6-11 Entering Activity will appear.
6 - 6
CHAMBER INITIALIZATION
March 15
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......