![Capintec CRC-25R Owner'S Manual Download Page 207](http://html.mh-extra.com/html/capintec/crc-25r/crc-25r_owners-manual_494176207.webp)
CAPINTEC, INC
CRC
®
-25R
Uncertainty Due to
Radioisotopes
Calibration
Setting
Number
Syringe
Corr. %
Published
Data %
Half-Life
(NCRP-58)
Ref. Comments
204
Tl Thallium
420
×
100
2
2
3.78 Y 2
NDT70
207
Bi Bismuth
846
1.7 32.2
Y
9
NBS73
Ref. for 1064, 569.7,
76.7, & 1772 keV.
208
Tl Thallium
571
÷
2
3.053 M 4
NM75
212
Pb
Lead
101
10.64 H 1
NM75
Decays to
212
Bi. Eqb.
after 1 hr. See App. IV.
212
Bi Bismuth
489
×
10
60.55 M 6
NM75
212
Pb Lead
(Eqb.
212
Bi)
158
Reading
gives
212
Pb
Act. in eqb. sample.
Eqb. after 8 hours.
212
Bi Bismuth
(Eqb.
212
Pb)
135
Reading
gives
212
Bi
Act. in eqb. sample.
212
Pb
212
Bi
Lead
Eqb.
Bismuth
030 or
146
×
2
Reading gives sum of
212Pb & 212Bi activity
in equilibrium sample.
224
Ra Radium
646
×
100
3.66 D 4
NM75
226
Ra Radium
+ chain of
daughters
778
0.5
1600 Y 7
NBS73
Reading in grams.
Commonly referred to
as "Radium "Source.
1.025 g/Ci of Ra-226.
239
Np
Neptunium
147
6
2.355 D 4
ORNL76
241
Am
Americium
055
4
1
432.2 Y 5
LMR69
Ref. for 59.5 & 14 keV.
June 09
APPENDIX II
A2 - 15
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......