CAPINTEC, INC
CRC
®
-25R
Place TOP of
Strip B in Chamber
Any Key to Continue
Figure 11-41 2 Strips Measurement Message – Top B
Place the Top of Strip B in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-39 QC Test Measurement Screen will re-appear.
Press
ENTER
to accept the measurement. Figure 11-42 2 Strips Measurement Message –
Place BOTTOM of
Strip B in Chamber
Any Key to Continue
Figure 11-42 2 Strips Measurement Message – Bottom B
Place the Bottom of Strip B in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-39 QC Test Measurement Screen will re-appear.
Press
ENTER
to accept the measurement. Figure 11-43 2 Strips Results Screen will appear.
QC TEST Two Strips
Free Tc99m
xx.x%
Reduced/Hydrolized
xx.x%
Purity
xx.x%
Any Key to Continue
Figure 11-43 2 Strips Results Screen
Press any key (except
HOME
) to continue. If a printer is attached to the system, a QC Test
2 Strip report will print. The report will show the measured values and results.
11 - 18
ENHANCED TESTS
August 07
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......