CAPINTEC, INC.
CRC
®
-25R
Moly Assay Limit
The Moly Assay default limit is set to 0.150 µCi/mCi (or 0.150 MBq/GBq). The user may
change this value.
Select
MO/TC LIMIT
from Figure 6-19 Moly Setup Menu. The current limit is displayed:
Mo/Tc LIMIT:
0.150 µCi/mCi
OK? Y or N
Figure 6-22 Mo/Tc Limit Verification Screen
If the displayed value is acceptable, press
YES
to keep the value and return to Figure 6-19
Moly Setup Menu.
If the displayed value is not acceptable, press
NO
. Figure 6-23 Mo/Tc Limit Entry Screen will
appear. The value is input in µCi/mCi (or MBq/GBq if in Becquerel mode):
INPUT
Mo/Tc Limit
in µCi/mCi
_
Figure 6-23 Mo/Tc Limit Entry Screen
Input the new value and press
ENTER
. Figure 6-22 Mo/Tc Limit Verification Screen will re-
appear displaying the new value.
Note: The minimum value that can be input is 0.001 µCi/mCi (0.001 MBq/GBq). The
maximum value that can be input is 9.99 µCi/mCi (9.99 MBq/GBq).
If the displayed value is acceptable, press
YES
to keep the value and return to Figure 6-19
March 15
CHAMBER INITIALIZATION
6 - 13
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......