CAPINTEC, INC
CRC
®
-25R
Place BOTTOM of
Strip B in Chamber
Any Key to Continue
Figure 11-49 HMPAO Measurement Message – Bottom B
Place the Bottom of Strip B in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-46 QC Test Measurement Screen will re-appear.
Press
ENTER
to accept the measurement. Figure 11-50 HMPAO Measurement Message –
Top C screen will appear:
Place TOP of
Strip C in Chamber
Any Key to Continue
Figure 11-50 HMPAO Measurement Message – Top C
Place the Top of Strip C in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-46 QC Test Measurement Screen will re-appear.
Press
ENTER
to accept the measurement. Figure 11-51 HMPAO Measurement Message –
Bottom C screen will appear:
Place BOTTOM of
Strip C in Chamber
Any Key to Continue
Figure 11-51 HMPAO Measurement Message – Bottom C
Place the Bottom of Strip C in the Chamber and press any key (except
HOME
) to perform
the measurement. Figure 11-46 QC Test Measurement Screen will re-appear.
August 07
ENHANCED TESTS
11 - 21
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......