![Capintec CRC-25R Owner'S Manual Download Page 121](http://html.mh-extra.com/html/capintec/crc-25r/crc-25r_owners-manual_494176121.webp)
CAPINTEC, INC.
CRC
®
-25R
Input the User ID number (reference the USER ID section on page 10-8) and press
ENTER
to acknowledge the usage message.
If the calculated ratio is within limits and is useable for more than 12 hours, Figure 10-19
Mo99/Tc99m with Volume Entry screen will appear.
Mo99/Tc99m:
0.058 µCi/mCi
VOLUME? ml
_
Figure 10-19 Mo99/Tc99m with Volume Entry
Input the volume of the elution in ml and press
ENTER
. The concentration of the elution is
calculated and displayed as shown in Figure 10-20 Elution Concentration screen.
Note: The minimum value that can be input is 1.000. The maximum value that can be input
is 100.0.
Tc99m:
1.232 Ci
123.2mCi/ml
A n y K e y t o C o n t i n u e
Figure 10-20 Elution Concentration
Press any key (except
HOME
) to print (if a printer is attached to the system) a ticket with the
details of the assay.
The prompt “PRINT AGAIN” will appear. If
YES
is pressed, another ticket is printed. If
NO
is
pressed, the
Tests Menu
will appear.
March 15
MEASUREMENT PROCEDURES
10 - 15
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......