CAPINTEC, INC
CRC
®
-25R
Input the time as hhmm in 24 hour time format (example: 1:25 PM would be 1325, 9:15 AM
would be 0915) and press
ENTER
. The time is checked for validity. If the input time is not
valid, a beep will sound and “TIME ERROR” will be displayed. The time must be re-entered.
The input activity, the date and time of use, and the volume required for that activity will be
displayed. The prompt “OK? Y or N” will appear. Press
NO
to return to the Figure 12-22
Required Activity Entry Screen.
If the displayed information is correct, withdraw the displayed volume and press
YES
. Figure
12-25 Measure Activity Screen will appear.
MEASURE
ACTIVITY
Any Key to Continue
Figure 12-25 Measure Activity Screen
Place the sample in the Chamber and press any key to continue. Figure 12-26 Withdraw
Activity Measurement Screen will appear.
Tc99m
ENTER to Accept
Figure 12-26 Withdraw Activity Measurement Screen
When the activity has stabilized, press
ENTER
to accept the measurement. The Inventory
item will update with the new volume and activity and Mo to Tc ratio for Tc99m (unless 0 was
entered for the ratio). Figure 12-2 Inventory Menu will re-appear.
PRINTING THE INVENTORY
If a printer is attached to the system, the Inventory can be printed.
To print the Inventory, select
from Figure 12-2 Inventory Menu.
12 - 12
INVENTORY
August 07
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......