CAPINTEC, INC
CRC
®
-25R
Tc99m
6.27mCi
ENTER to Accept
Figure 11-46 QC Test Measurement Screen
Press
ENTER
to accept the measurement. Figure 11-47 HMPAO Measurement Message –
Place BOTTOM of
Strip A in Chamber
Any Key to Continue
Figure 11-47 HMPAO Measurement Message – Bottom A
Place the Bottom of Strip A in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-46 QC Test Measurement Screen will re-appear.
Press
ENTER
to accept the measurement. Figure 11-48 HMPAO Measurement Message –
Top B screen will appear.
Place TOP of
Strip B in Chamber
Any Key to Continue
Figure 11-48 HMPAO Measurement Message – Top B
Place the Top of Strip B in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-46 QC Test Measurement Screen will re-appear.
Press
ENTER
to accept the measurement. Figure 11-49 HMPAO Measurement Message –
11 - 20
ENHANCED TESTS
August 07