CAPINTEC, INC.
CRC
®
-25R
STANDARD
LINEARITY TEST
TIME FOR
MEASUREMENT 2
IN HOURS
Figure 6-45 Standard Test – Test Time Entry
Input the time (in hours) and press
ENTER
. The above screen will be repeated for each
measurement.
After all the times are input, the Standard Test Verification screen will appear. The example
shown (Figure 6-46 Standard Test Verification) is for 12 tests at 1 hour increments:
HOURS FOR TESTS
1: 0 2: 1 3: 2
4: 3 5: 4 6: 5
7: 6 8: 7 9: 8
10: 9 11: 10 12: 11
OK? Y or N
Figure 6-46 Standard Test Verification
If the displayed times are incorrect, press
NO
. Figure 6-44 Number of Tests Entry will re-
appear.
If the displayed times are correct, press
YES
. Figure 6-1 Other Menu will reappear.
Setting Up Lineator Test
This test requires 8 measurements. A Lineator Test Calibration procedure must be
performed before any Lineator tests.
Select
LINEATOR
from Figure 6-43 Linearity Test Selection Menu. Figure 6-47 Lineator
Calibration Prompt screen will appear. The example below is given for the 5
th
measurement.
March 15
CHAMBER INITIALIZATION
6 - 25
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......