CAPINTEC, INC
CRC
®
-25R
Measured x.xxmCi
OK? Y or N
Figure 11-16 Standard Linearity Test First Measurement Verification
To repeat the measurement or if the measurement is not acceptable, press
NO
.
Figure 11-14 Standard Linearity Test First Measurement Prompt will re-appear.
If the measurement is acceptable, press
YES
. If a printer is attached to the system, a
Standard Linearity Test First Measurement report will print and the prompt to print
again will appear.
To print the test results again, press
YES
. To return to Figure 11-1 Tests Menu,
NO
.
The report will show the predicted and actual activity and the variance.
Second Through Last Measurements
Perform the measurements at time intervals close to that entered when the test was
defined. The actual elapsed time will be used in the calculations.
From Figure 11-1 Tests Menu, select
ENHANCED
. Figure 11-2 Enhanced Tests
Menu will appear.
Select
LINEARITY
. Figure 11-17 Standard Linearity Test – Previous Test Results
Screen will appear.
In the example, 2 measurements have been made.
STANDARD LINEARITY
# MEASURED %VAR
1 x.xxmCi x.xx
2 x.xxmCi x.xx
New Test? Y or N
Figure 11-17 Standard Linearity Test – Previous Test Results Screen
11 - 8
ENHANCED TESTS
August 07
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......