Standard Test
The Standard Test measures the activity of Tc99m over a period of time. When sufficient
measurements have been made, a least-squares fit is performed and the deviation of the
results from this fit are reported.
First Measurement
Figure 11-14 Standard Linearity Test First Measurement Prompt will appear for the
first measurement.
Test # 1
First Measurement
Any Key to Continue
Figure 11-14 Standard Linearity Test First Measurement Prompt
Place the linearity source in the Chamber and press any key (except
HOME
) to
perform the measurement. Figure 11-15 Standard Test Measurement Screen will
appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-15 Standard Test Measurement Screen
ENTER
to accept the measurement. Figure 11-16 Standard Linearity Test First
Measurement Verification will appear.
August 07
ENHANCED TESTS
11 - 7
Summary of Contents for CRC-25R
Page 2: ......
Page 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Page 227: ......
Page 228: ......