9
5.
Installation
The unit should be placed on a sturdy bench. Although the unit has a very high static
mass to vibrating mass ratio and absorbent rubber feet, a rigid support bench will
enhance immunity from secondary vibration transmission.
If any accessories
– microscope/magnifier attachment or light source - have been
supplied with the instrument, reference should be made to the appropriate section of
this manual for the fitting and setting up procedure.
Before connecting the unit to a mains supply, the unit must be set for your particular
voltage supply.
The voltage is set by prising out the fuse holder drawer and re-inserting it such that
the voltage legend for your supply is aligned with the mark on the inlet moulding. See
below.
Voltage supply selector (example shown is set to 220-240V)
The inlet moulding accepts a standard IEC socket. Where possible a standard mains
lead - IEC socket/mains plug
– suitable for your mains outlet will have been supplied
with the instrument. The instrument must not be operated unless it is connected to a
suitably earthed (grounded) mains supply.
See Section 7 for fitting the light source, magnifier and microscope.
On/Off switch
Mains inlet
Voltage selector &
fuse drawer
Voltage
alignment marks
Summary of Contents for 7000smz
Page 26: ...26 Correct any problems and press SLICE to retry...
Page 35: ...35...