When the
‘Opti-Cal’ calibration device is plugged into the instrument its presence will
automatically be detected and it will take control of the amplitude and frequency
settings, adjusting them to a magnitude that allows best adjustment of the blade
alignment error. The alignment process also automatically moves the blade so that
the cutting edge is in its optimal position relative to the calibration device for best
alignment evaluation.
To enter the alignment routine, press the
‘LOAD BATH’ key (and hold for 2 seconds).
The bath table will be moved to its lowest position to allow the
‘Opti-Cal’ to be loaded.
Press
‘AUTO REPEAT’ to begin the blade alignment routine.
The alignment routine can also be accessed from the main menu.
When the unit has been connected,
the
instrument
will
detect
the
connection and the display will change
as shown.
Install the alignment device and new
blade.
Press
‘SLICE’ to continue.