13
6.3
Load or Unload the specimen bath
Press the
‘LOAD BATH’ key (and hold
for 2 seconds) and the display will
change to that shown. The bath table
will be moved to its lowest position to
allow a specimen bath to be loaded (or
an existing one removed).
Once the table has reached its lowest
position the display will be as shown.
The specimen bath is located on the
rising table in a sprung loaded dovetail
mount.
To remove the bath, move the clamp
lever to the left and pull the bath away
from the instrument.
Pressing
‘LOAD BATH’ keypad will
cause the table to rise to a preset
(nominal) position.
The display will then revert to the
‘home’ display.
Pressing
‘Auto Repeat’ keypad will
enter the blade alignment routine. See
Section 6.9
‘Alignment of the blade
with the axis of oscillation
’ for detailed
guidance on the use of the
‘Opti-Cal’
alignment calibrator.
Note: users wanting to
customise the tissue mount
must ensure that the total
weight of ice bath, tissue
bath and contents must not
exceed 800g, or this would
cause over-load of the
motor.