![Campbell SkyVUE 8 Product Manual Download Page 87](http://html1.mh-extra.com/html/campbell/skyvue-8/skyvue-8_product-manual_3808856087.webp)
h5
(4
characters)
=
Height
of
the
5th
cloud
layer
in
10s
of
metres
or
100s
of
feet,
if
no
5th
layer
is
reported
h5
=
////.
CR
LF =
Carriage
Return
+
Line
Feed
NOTE:
Cloud amounts and heights cannot be reported until the SkyVUE 8 has been operating for 30
minutes.
LINE
4
scale
V
res
V
n
V
energy
V
lt
V
tr
V
ti_bl
V
pulse
V
rate
V
sum
CR
LF
where
scale
(5
characters)
=
Scale
parameter,%,
0
to
99999.
100%
is
default
res
(2
characters)
=
Backscatter
profile
resolution
in
metres
n
(4
characters)
=
Profile
length
energy
(3
characters)
=
Laser
pulse
energy,
%.
lt
(3
characters
including
leading
+/-)
=
Laser
temperature,
degrees
C
ti
(2
characters)
=
Total
tilt
angle,
degrees
bl
(4
characters)
=
Background
light,
millivolts
at
internal
ADC
input
(0
to
2500)
pulse
(4
characters)
=
Pulse
quantity
x
1000
(0000
to
9999)
rate
(2
characters)
=
Sample
rate,
MHz,
(00
to
99)
sum
=
(3
characters)
Sum
of
detected
and
normalized
backscatter,
0
to
999.
Multiplied
by
scaling
factor
times
10
4
.
At
scaling
factor
100
the
SUM
range
0
to
999
corresponds
to
integrated
backscatter
0
to
0.
srad
-1
.
CR
LF
=
Carriage
Return
+
Line
Feed
LINE
5
SSSSSSSSSSSSSSSSSSSSSSSSSSSSS........(2048 x 5 bytes) CR LF
The two-way attenuated normalized backscatter profile (
CR LF = Carriage Line Feed
LINE 6
ETX CRC-16 EOT CR LF