![Campbell SkyVUE 8 Product Manual Download Page 84](http://html1.mh-extra.com/html/campbell/skyvue-8/skyvue-8_product-manual_3808856084.webp)
h2
(4
characters)
=
Height
of
the
2nd
cloud
layer
in
10
s
of
metres
or
100
s
of
feet,
if
no
2nd
layer
is
reported
h2
=
////.
d3
(1
character)
=
Cloud
amount
of
the
3rd
layer
in
oktas
h3
(4
characters)
=
Height
of
the
3rd
cloud
layer
in
10
s
of
metres
or
100
s
of
feet,
if
no
3rd
layer
is
reported
h3
=
////.
d4
(1
character)
=
Cloud
amount
of
the
4th
layer
in
oktas
h4
(4
characters)
=
Height
of
the
4th
cloud
layer
in
10
s
of
metres
or
100
s
of
feet,
if
no
4th
layer
is
reported
h4
=
////.
d5
(1
character)
=
Cloud
amount
of
the
5th
layer
in
oktas
h5
(4
characters)
=
Height
of
the
5th
cloud
layer
in
10
s
of
metres
or
100
s
of
feet,
if
no
5th
layer
is
reported
h5
=
////.
CR
LF
=
Carriage
Return
+
Line
Feed
NOTE:
Cloud amounts and heights cannot be reported until the SkyVUE 8 has been in operation for
30 minutes.
LINE 4
ETX CRC-16 EOT CR LF
where
ETX = End-of-Text character
CRC-16 (4 characters) = CRC-16 Checksum
EOT = End-of-Transmission character
CR LF = Carriage Line Feed
6.3.4 MESSAGE 004 (profile, sky condition) — default
message
Example message line outputs
CS0001004
10 092 00698 ///// ///// ///// 800000000000
99 //// 0 //// 0 //// 0 //// 0 ////
00100 05 2048 100 +40 02 0074 0070 30 000
071fc024d………………….000 (Line output cropped for clarity)
6212