![Campbell SkyVUE 8 Product Manual Download Page 78](http://html1.mh-extra.com/html/campbell/skyvue-8/skyvue-8_product-manual_3808856078.webp)
Table 6-2: Middle alarm word for CS messages
Bit
1
Description
XXXX 0200 XXXX
DSP factory calibration stored in flash has failed its signature check
XXXX 0100 XXXX
DSP board OS signature test failed
XXXX 0080 XXXX
DSP board RAM test failed
XXXX 0040 XXXX
DSP boards on board PSUs are out of bounds
XXXX 0020 XXXX
TOP board non-volatile storage is corrupt
XXXX 0010 XXXX
TOP board OS signature test has failed
XXXX 0008 XXXX
TOP boards ADC and DAC are not within specifications
XXXX 0004 XXXX
TOP boards on board PSUs are out of bounds
XXXX 0002 XXXX
Communications have failed between TOP board and the DSP
XXXX 0001 XXXX
Photo diode background radiance is out of range
1
Alarm words are sum of error bits.
Table 6-3: Least significant alarm word for CS messages
Bit
1
Description
XXXX XXXX 8000
Photo diode temperature is out of range
XXXX XXXX 4000
Photo diode is saturated
XXXX XXXX 2000
Photo diode calibrator temperature is out of range
XXXX XXXX 1000
Photo diode calibrator has failed
XXXX XXXX 0800
The sensor could not reach the desired gain levels
XXXX XXXX 0400
Laser run time or maximum laser drive current has been exceeded
XXXX XXXX 0200
Laser temperature out of range
XXXX XXXX 0100
Laser thermistor failure
XXXX XXXX 0080
Laser is obscured. This can only be set if the laser is on
XXXX XXXX 0040
Laser did not achieve significant output power
XXXX XXXX 0020
Laser max power exceeded
XXXX XXXX 0010
Laser max drive current exceeded