![Campbell SkyVUE 8 Product Manual Download Page 17](http://html1.mh-extra.com/html/campbell/skyvue-8/skyvue-8_product-manual_3808856017.webp)
WARNING:
Removal of electronic module covers or connectors while the unit is powered will expose the
operator to potentially hazardous voltages and risk damage to the sensor.
The SkyVUE 8 has electrical and laser warning labels on the exterior and interior of the unit.
It is fitted with a hex-key access panel. It is recommended that the hood and access panel door
are not opened in conditions of rain, hail or snow.
Isolate the sensor before removing internal components, including the cover to the avalanche
photodiode (APD) module. Only trained personnel should disassemble the instrument.
Mains connectors are shrouded to prevent touching of the contacts. The mains supply should be
isolated when connecting and disconnecting the cables to the sensor.
Where an isolator switch is fitted, this shall be a two-pole isolator, located as near to the sensor
as possible.
The SkyVUE 8 must be properly grounded by a licensed and qualified electrician to protect
against voltage leakage shock risk (
Campbell Scientific recommends that RCD protection units be used with all sensors. See
(p. 20) for further information.
Mains powered heaters are enclosed to prevent contact.
CAUTION:
When powered, the heaters may operate automatically and without warning. They may
remain hot when not powered. Follow the isolation precautions, to avoid shock and burn
hazards.
The sealed battery has cables and protected connectors to prevent shorts. Avoid shorting the
battery to protect it from damage and to avoid burns to personnel through contact with hot
surfaces.
2. Product overview