CRVW3 3-Channel Vibrating-Wire Datalogger
98
6.
Set the
Run_Diag
value to 1 to trigger the measurement. Watch and wait
for the value to be set back to zero (0) to indicate that the test measurement
is complete. Ensure
Diag_Result_Code
is zero (0).
7.
Open the
File Control
screen using the
File Control
button.
8.
Note the
Spectrum.bin
and
TimeSeries.bin
files located on the CPU drive.
The date and time on these files should match the time of the test
measurement. (A timestamp of zero on the files indicates that they do not
yet exist, the test has not yet been performed).
9.
Select
only
the file
TimeSeries.bin
and press the
Retrieve…
button.
10.
Save the file to a known folder location on your
LoggerNet
computer
station.
11.
Open
DevConfig
, choose
CRVW Series
for the device, select the
Off
Line Analysis
tab. Press the
Read File
button, and browse to the
TimeSeries.bin
file.
12.
The contents of the file will be processed, and
DevConfig
will display the
spectrum and time-series information as recorded by your remote CRVW3
Summary of Contents for CRVW3