4.3
Corrosion Resistance
Standard housing versions feature a double-layer powder coat for enhanced corrosion protection,
designed to be highly resistant to decontamination agents commonly used in Life Science
and Clean Process environments.
Camfil has chosen to evaluate corrosion resistance of its metal products with respect to
decontamination products most commonly used by our customers.
•
Hydrogen Peroxide (H
2
O
2
)
https://en.wikipedia.org/wiki/Hydrogen_peroxide
•
Peracetic Acid (CH
3
CO
2
OH)
https://en.wikipedia.org/wiki/Peracetic_acid
•
Actril
TM
Cold Sterilant
http://www.medivators.com/products-and-services/renal-systems/cold-sterilant
s
➔
A comprehensive testing report is available, upon request
➔
Corrosion resistance to other decontaminants can be determined in our laboratory, upon request
PLEASE NOTE:
CamSafe is designed for indoor use in Life Science applications with HVAC processed air.
No corrosion resistance class according to ISO 12944 is established.
Store parts in a dry area protected from weather, prior to installation.
For heavy duty applications in Life Science and/or environments with high condensation like :
•
High moisture / low tempe severe decontamination
(heavy acid or case concentrations)
•
Negative temperature warehouses for bulk vaccine storage
•
Blood fractioning applications
we recommend using stainless steel versions.
For these or similar cases, we can supply CamSafe housings in stainless-steel AISI 304L or 316L,
depending on application severity.