Install & Operating Manual
CamSafe
Filter Validation with MIP (continued)
Start with the probe fully inserted
and then move over the filter
with a slow pulling motion
Please Note : zero count
in the middle of the filter
CAUTION
Pull in a regular motion.
The probe and hoses should not
suffer any jolts during the
procedure. This could lead to
incorrect readings.
If an increased particle count is
detected, stop the probe
at that point.
If necessary, the leak may be
localized by moving the probe
back and forth.
CAUTION
Switch off the
particle counter before
disconnecting as the port
seals airtight automatically
on both fittings!
After completing the measuring
procedure, remove measuring
hoses from the self-sealing
connectors.
Push back the mobile downstream
probe fully and screw on the
protection cap.