![Camfil CamSafe Installation & Operating Manual Download Page 40](http://html.mh-extra.com/html/camfil/camsafe/camsafe_installation-and-operating-manual_487851040.webp)
EMC_SPEC018_C2.doc
Rev. D
–
2020 05 19
40/44
Install & Operating Manual
CamSafe
11.1
Filter Validation with MIP
Before testing :
•
Perform a zero-count
•
Check airtightness of the door
•
Check airtightness of the filter gasket seat
Connect
Downstream Sampling Port
(mobile)
Connect
Upstream Concentration Port
(static)
CAUTION
For upstream concentration
measurement, use a dilution
system
CAUTION
The extended downstream probe
must not be subjected
to horizontal or vertical load
to avoid damage
PLEASE NOTE:
Connect a particle counter or photometer using G1/4 fittings
RIEGLER 246.32 or 246.12 or equivalent.
Switch off the particle counter before disconnecting as the port seals airtight
automatically on both fittings!
CAUTION
Do not use a MIP system with a bent or otherwise damaged mobile probe
as this might result in leakage through the probe bulkhead.
Replacement of the probe involves breaking the containment barrier of the housing !
Use original Camfil parts and specialist staff only.