89
MC-III™ EXP Flow Analyzer
Section 6
Section 6 - MC-III EXP Maintenance
The MC-III™ EXP is designed to provide many years of service with minimal maintenance. Batteries require
periodic replacement, and battery life depends on whether battery power is the primary or secondary power
source.
All configuration settings are stored in nonvolatile memory; therefore, configuration settings will not be lost
in the event of battery failure.
The circuit assembly or keypad may also require replacement over the life of the instrument. Procedures are
provided in this section.
Lithium Battery Replacement
The MC-III EXP uses a lithium battery with a life expectancy of approximately 2 years. Due to the flat
discharge characteristics of the lithium battery, it is difficult to determine how much life remains in a battery
at any given time. To preserve configuration and accumulated volume data, replace the battery at 2-year
intervals.
!
WARNING: CSA-APPROVED DEVICE. To prevent ignition of hazardous atmospheres, do not
remove the cover unless the area is void of combustible gas and vapors. Replace the battery
pack only with battery part number 9A-100005111.
!
WARNING: ATEX-APPROVED DEVICE. To prevent ignition of hazardous atmospheres, do not
remove the cover unless the area is void of combustible gas and vapors. Replace the battery
pack only with battery part number 9A-90099000.
!
WARNING: The lithium battery that powers the MC-III EXP is a sealed unit; however, should a
lithium battery develop a leak, toxic fumes could escape upon opening the enclosure. Ensure
that the instrument is in a well-ventilated area before opening the enclosure to avoid breathing
fumes trapped inside the enclosure. Exercise caution in handling and disposing of spent or
damaged batteries. See additional information on lithium batteries in
this manual.
Important: Press the ENTER/SAVE key on the keypad before replacing the lithium battery to save
the accumulated total to nonvolatile memory. Once the battery is replaced and power is
restored to the unit, the last saved accumulated total will be displayed in the LCD.
The lithium battery is secured inside the enclosure by a velcro strap and connected to a connector (J1) near the
top of the circuit assembly.
To replace a lithium battery in the MC-III EXP, perform the following steps:
1. Unscrew the cover of the enclosure counter-clockwise until it separates from the main body of the enclo-
sure.
2.
Using a small standard blade screwdriver, remove the two #4-40 × 7/8” screws located to the right and
).
Summary of Contents for NUFLO MC-III EXP
Page 1: ...NUFLO TM MC III EXP Flow Analyzer User Manual Manual No 9A 50165003 Rev 06 Valves Measurement ...
Page 18: ...12 Section 1 MC III EXP Flow Analyzer ...
Page 35: ...29 MC III EXP Flow Analyzer Section 2 ...
Page 46: ...40 Section 3 MC III EXP Flow Analyzer ...
Page 81: ...75 MC III EXP Flow Analyzer Section 4 ...
Page 93: ...87 MC III EXP Flow Analyzer Section 5 Figure 5 17 Dropdown menu for exporting log data ...
Page 94: ...88 Section 5 MC III EXP Flow Analyzer ...
Page 105: ...99 MC III EXP Flow Analyzer Section 6 ...
Page 114: ...A 9 MC III EXP Flow Analyzer Appendix A ...
Page 121: ...B 7 MC III EXP Flow Analyzer Appendix B ...
Page 123: ...C 2 Appendix C MC III EXP Flow Analyzer ...
Page 141: ...D 18 Appendix D MC III EXP Flow Analyzer ...
Page 143: ......