Troubleshooting 112
P/N 113564 Rev. 2
Twister II Robot User’s Manual
Caliper Life Sciences, Inc.
The instrument was moved
Re-teach the instrument positions (see
The wrong rack is in a rack position
(e.g. Rack 3 is in rack position 2)
Either place the proper rack in the position or
re-teach the storage positions (see
).
The plates are rubbing on the input
or output racks.
Re-teach the storage positions (see
).
The Twister II is not lowering the
plates all the way down before
releasing them.
Re-teach the storage positions (see
).
Make sure the storage configuration is correct
if using the Stack Storage resource for CLARA.
Issue 3
The Twister II does not properly place materials into an instrument or the
storage racks.
Possible Cause
Solution