26 - ENG PN MN239-C4 E | User Manual
NewLife Family
Guidance and Manufacturer’s Declaration ± Electromagnetic Immunity
The NewLife is intended for use in the electromagnetic environment specified below. The customer or the user
of the NewLife should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment ±
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF
communications equipment should
be used no closer to any part of the
NewLife, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
from 80 MHz to 800 MHz
from 800 MHz to .5 GHz
where
P
is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and
d
is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the NewLife Family is used
exceeds the applicable RF compliance level above, the NewLife should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating
the NewLife Family.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
d
=
1.2
√
P
d
=
1.2
√
P
d
=
1.2
√
P
Summary of Contents for NewLife Series
Page 290: ...NewLife Family 290 ELL MN239 C4 E SOC 65 NewLife...
Page 291: ...NewLife Family MN239 C4 E ELL 291 NewLife NewLife 1 1 1 A I B lpm E H 1 1 NewLife...
Page 292: ...NewLife Family 292 ELL MN239 C4 E 3 1 2 CAIRE NewLife NEWLIFE INTENSITY...
Page 293: ...NewLife Family MN239 C4 E ELL 293 I PINE SOL LESTOIL...
Page 294: ...NewLife Family 294 ELL MN239 C4 E NEWLIFE 30cm 1ft GFCI LPM 2LPM 1 8LPM 2LPM...
Page 298: ...NewLife Family 298 ELL MN239 C4 E 1 2 I Elite Intensity LPM 2LPM 1 8LPM 2LPM...
Page 299: ...NewLife Family MN239 C4 E ELL 299 NewLife NewLife NewLife 10 micron 1 1 4 2 3 4 6 5...
Page 304: ...NewLife Family 304 ELL MN239 C4 E 1 6 6 2 1 2 3 4 1 2 1 T 2 3 1 3 30 7 4 5 6 7...
Page 306: ...NewLife Family 306 ELL MN239 C4 E PINE SOL LESTOIL 1 2 3...
Page 307: ...NewLife Family MN239 C4 E ELL 307 10 LPM 30 cm 12 in NewLife...
Page 308: ...NewLife Family 308 ELL MN239 C4 E 30 5 cm 12 in I NewLife I...
Page 418: ...418 ENG PN MN239 C4 E User Manual NewLife Family NOTES...
Page 419: ...PN MN239 C4 E User Manual ENG 419 NewLife Family NOTES...